59
Mahr GmbH, Millimar C1208/C1216/C1240
M1240 Millimar automatically sends the
measured values or data set-
tings requested by a connected
PC and ceases data transfer with
<CR>.
MARTALK Enables data and settings to be
queried by a connected PC and
Millimar configuration to be
changed using the configuration
program D 1000 X or D 1000 S.
i
As long as the connected Millimar
is being configured, the D 1000 S
program automatically switches
the Millimar’s interface protocol to
MarTalk by sending a break signal
to the Millimar on the reception
line. As soon as the configuration is
complete, the D 1000 S reinstates
the interface protocol selected by
the user.
When the connected PC or pro-
grammable controller is switched
on or off, a break signal can be
unintentionally triggered and the
interface protocol “MarTalk” acti-
vated. However, if Millimar does not
receive a valid MarTalk command
within the following 2 seconds, the
interface protocol set on Millimar by
the user will be automatically rein-
stated.
When using a terminal emula-
tor (e.g. Hyperterm) a break signal
can be triggered by pressing the
key combination CTRL + PAUSE/
BREAK. The “DTR/DSR” line is then
deactivated using the pull-down
menu “CALL --> SEPARAT.”