33
Mahr GmbH, Millimar C1208/C1216/C1240
MEANVAL The mean value of all the
individual measured values
is continuously displayed
through out the measure-
ment period. The mean value
is calculated according to the
follo wing formula:
(Measured value 1 + Mea-
sured value 2 + ..... + Mea-
sured value x) ÷ Number x of
in dividual measured values.
6. Press START once the desired function (the
desired feature) is displayed. The function
now flashes.
7. Press START again to adopt the flashing set-
ting (feature). The display then returns to the
catalog of functions and parameters item
FNCTION (1.2)”.
If at this point either the MASTER or DATA
key is pressed instead of START, the display
returns to the feature selection list (cf. point
5 of this section). A different function (a dif-
ferent feature) can then be selected.
8. If only one feature is indicated on the display,
press START to exit the catalog of functions
and parameters. If, however, two features are
displayed, follow the instructions in points 9
to 17 of this section.
9. Press the ESC key. The setting “FEAT. 1” is
displayed.
10. Press the MASTER key. The setting “FEAT.
2” is displayed (“2 FEAT.” must have been
set under “SETTING --> DISPLAY -->
FEATURE (4.1.1)”).
11. Press the MENU key again. “FACTOR (2.1)”
is displayed.
12. Use the DATA and MASTER keys to select
FNCTION (2.2)”.
13. Press the MENU key. The current feature
flashes.
14. Use the DATA and MASTER keys to se-
lect the feature to be calculated. The choice
of settings listed in point 5 are available to
choose from.
15. Press START once the desired function (the
desired feature) is displayed. The function
now flashes.
16. Press START again to adopt the flashing set-
ting (feature). The display then returns to the
catalog of functions and parameters item
FNCTION (2.2)”.
17. If no further settings are to be made in the
catalog of functions and parameters, press
the START key. The standard display ele-
ments appear. The selected feature’s desig-
nation is displayed in front of the measuring
result: MAX, MIN, MAX+MIN, MAX-MIN, M.
The “M” here stands for mean value.
If further settings are necessary, navigate
to the appropriate point of the catalog of
functions and parameters using the ESC,
MASTER, DATA and MENU keys and carry
out the settings as required.
i
If the function/feature has to be
changed frequently, this function
from the catalog of functions and
parameters can be assigned to the
SELECT key to enable rapid access.