Activating the measured value memory
function
Select “AUTOM (3.1.1.2)” under PROCESS -->
MEASURE --> MEAS.MOD. The letter “T” ap-
pears in the footer of the display, indicating that
the measured value memory function is active.
Controlling the measurement period for
the measured value memory function
1. With the START key
When the START key is pressed for the first time,
any data saved in the measured value memo-
ries is deleted and the measurement period is
started. The data stored in the measured value
memories is continuously updated until the
START key is pressed again. The measurement
period ends and the desired feature is calculated
from the saved data and displayed.
2. By entering a specified period of time
The measurement period during which data in
the measured value memory is updated is speci-
fied in seconds. The measurement period can be
set under
PROCESS -> MEASURE ->TIMER (3.1.2)
between 0.1 seconds and 999.9 seconds.
Measurement is started either by pressing the
START key or by an appropriate signal at the
control input. At the end of the measurement
period, measurement ceases, i.e. the data in the
measured value memories is no longer updated.
3. Using a control input
The measured value storage can also be acti-
vated and deactivated using control signals. This
approach is suitable when e.g. the measured
value memory is to be activated and deactivated
using contacts that are fitted to the measuring
instrument.