Pellicon-2 and 3 Mini Holder User Guide 13
Setup for Autoclaving
Remove all filters and filter gaskets from the Pellicon 2 and 3 Mini Holder before 1.
autoclaving.
Rest the holder FEED and RETENTATE fittings and the end plate on a soft 2.
material to protect the fitting faces and end plate from being scratched in the
autoclave.
Do not autoclave with the holder in an upright position, because the feet will
compress and become loose.
Autoclave the Pellicon 2 and 3 Mini Holder at 121º C to 125º C for 30 minutes per 3.
cycle.