16 Probes and Biopsy
Operator’s Manual 16 - 23
Perform the following procedure:
1. Install the needle-guided bracket:
a. Put on the sterile probe sheath.
b. Open the retaining clamp, align the needle-guided bracket with the probe to align the
locating bulge on the needle guide with the locating grooves on the probe, then turn the
retaining clamp to align it with the probe.
c. When the retaining clamp is turned to the correct position, the locking nut will lock the
retaining clamp and the needle-guided bracket is then mounted in the correct position.
2. Remove the needle-guided bracket:
Hold the probe in your left hand. Unscrew the locking nut with your right hand to open the
retaining clamp, then raise the needle-guided bracket to separate the locating bulge from the
locating grooves.
4 Retaining clamp
5Probe
6 Locating groove