1-2 
non-affiliated to the manufacturer or by non-licensed personnel may affect 
the safety, performance and function of the product. 
  When used with electrosurgery equipment, the safety of patients should be 
ensured. 
  The patient's clinical condition and the working condition of the infusion 
pump must be monitored carefully, the alarm volume and alarm levels need 
to be set according to the actual needs. Operation and performance relying 
solely on the auditory alarm system alone is not sufficient, and setting the 
alarm at a low volume may endanger the patient. If the alarm volume is less 
than the surroundings volume, which can further lead to operators identify 
alarm mistakenly. 
  The interconnection of other infusion systems or ACCESSORIES to the 
PATIENT LINE may lead to performance degradation and failure to achieve 
the expected performance, the working condition of the pump and patient's 
clinical condition shall be monitored regularly. 
  Please carefully install the power line and cables with various accessories 
to prevent the patient from choking or suffocation caused by entanglement 
of the cables or by electrical disturbance. 
  The packaging materials must be disposed of in compliance with local 
laws and regulations or the hospital policy on waste management.They 
must be kept out of the reach of children. 
  Infusion tube knots, filter coagulation and occlusions arising from needle 
insertion can cause the pressure inside the infusion set to rise during 
infusion. When this occurs, removing the occlusion can cause excessive 
liquid to be infused into the patient, so appropriate measures should be 
taken. 
  The pump should not be placed more than 100cm above or below the level 
of the patient’s heart. The smaller the height difference between the pump, 
the more accurate the pressure test in the infusion cannula will be. 
  It is recommended that infusion pump with conventional infusion sets is 
used with infusion sets recommended by manufacturers (please refer to 
6.10.1 Infusion Pump with Conventional Infusion Sets for specific brands). 
When use of non-recommended infusion sets, please make sure to confirm 
relevant infusion performance (such as accuracy, air bubble and pressure) 
on infusion pump, and contact the company for calibration service. Air 
bubble calibration, Accuracy Calibration and Pressure Calibration shall be 
performed sequentially, the infusion sets can only be used after 
confirmation, otherwise Mindray Scientific is not responsible for infusion 
performance (such as accuracy, air bubble and pressure) and relevant 
alarm function of the infusion pump. Its accuracy cannot be guaranteed 
when the pump is used with an infusion set without calibration.