C-1
C Alarm Information
This chapter presents the alarm information of the infusionpump.Prompt information
for operation guidance will not be presented in this chapter.
The table shows the appropriate countermeasures for each piece of information
related to alarm triggering.If the problem still exists after operating according to the
countermeasures, please contact the company.
Alarm
Alarm
Reason Countermeasure
[Upstream
Occlusion]
High
When the pump is infusing, the
alarm is triggered due to an
occlusion in the infusion tube
between liquid source and
equipment. Infusion is stopped
after the alarm is triggered.
Press to cancel
alarm, check and eliminate
the source of the alarm.
[Downstream
Occlusion]
High
When the pump is infusing, the
alarm is triggered due to an
occlusion in the infusion tube
between the equipment and
the patient. Infusion is stopped
after the alarm is triggered.
Press to cancel
alarm, check and eliminate
the source of the alarm.
[Air in line] High
Size of single air detection
reaches to the preset value.
Infusion is stopped after the
Press to cancel
alarm, check and eliminate
[Accumulate air] High
Size of accumulate air bubble
accumulated in 1 hour reaches
to the preset value. Infusion is
stopped after the alarm is
triggered.
Press to cancel
alarm, check and eliminate
the air in line.
[Drop error] High
If drop sensor is installed
correctly and the switch of
[Drop sensor] is on, drop
sensor detects that the drop
rate deviates the preset value.
Infusion is stopped after the
Press to cancel
alarm, refill the drop
chamber correctly or place
the drop chamber
vertically or avoid the