EasyManua.ls Logo

Mindray BeneFusion VP5 - External Ports; Signal Output Interface; Specifications

Mindray BeneFusion VP5
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A-4
A.3.5 External Ports
Ports
AC power supply
port
One AC power supply port
Multifunction
interface
One multifunction interface with the following functions:
DC power supply port
RS232 interface
Nurse call interface
Drop sensor
interface (Optional)
One Drop sensor interface
A.3.6 Signal Output Interface
Nurse call signal output
Driving mode Relay drive
Electric
specification
≤60W, ≤2A, ≤36VDC, ≤25VAC
Isolation voltage >1500VAC
Action mode Normally open or normally closed (optional)
A.4 Specifications
Parameters Specifications Factory Default
Infusion set
standard
Infusion set used in conjunction with
infusion pump should meet the
requirements of ISO 8536-4:2004 Infusion
equipment for medical usePart 4:
Infusion sets for single use, gravity feed,
MOD
/
Compatible
conventional
set sizes (ml)
Infusion set diameter: 3.0-4.5mm
Infusion set thickness: 0.8-1.2mm
/
Rate
Unit of Rate (ml/h): 0.1-2000ml/h, the
minimum increment is 0.01ml/h
Unit of Rate (gtt/min): 1-(400*Drip /60)
gtt/min, the minimum increment is 1gtt/min
Note 1: The maximum rate of the infusion
system (infusion pump and infusion set) is
affected by inner and outer diameter,
material, elasticity of the infusion set and
---

Table of Contents

Related product manuals