A-5 
other factors. Therefore, infusion sets of 
different brand and model may differ in 
maximum rates. 
Note 2: The above declared rates are 
based on BOON A2 and B.Braun Intrafix 
Safeset infusion sets, and specified 
infusion sets,. temperature of 20±2
℃
Bolus Rate/ 
Purge Rate 
0.1-2000ml/h  / 
VTBI    0.1-9999.99ml 
--- 
Time  00:00:01-99:59:59 h:m:s  --- 
Volume  0.01-9999.99ml  0ml 
Select Mode 
Rate Mode, Body Weight Mode, Time 
Mode, Sequential Mode, Loading Dose 
Mode, Micro-infusion Mode and Ramp 
Rate Mode 
Pressure 
12 Levels are adjustable, respectively are 
(75, 150, 225, 300, 375, 450, 525, 600, 
750, 825, 900 and 975)mmHg 
For 75mmHg, the error is 
±50mmHg(6.7kPa); for other pressures, 
the error is ±20% or ±113mmHg (15.1kPa), 
whichever is larger. 
Maximum occlusion pressure is about 
1170mmHg. 
Note 1: The detected pressure of the 
infusion system (infusion pump and 
infusion set) is affected by inner and outer 
diameter, material, elasticity of the infusion 
set and other factors. Therefore, infusion 
sets of different brand and model may 
differ in the detected pressure range. 
Note 2: The above declared pressure is 
based on BOON A2 and B.Braun Intrafix 
Safeset infusion sets, and specified 
infusion sets,temperature of 20±2℃. 
450mmHg 
Air detection 
6 Levels are adjustable, respectively are 
(20, 50, 100, 250, 500, 800) μl 
Sensitivity of single air bubble is 20μl. 
Note 1: The detected air detection of the 
infusion system (infusion pump and 
infusion set) is affected by inner and outer 
diameter, material, elasticity of the infusion 
100μl