A-6
set and other factors. Therefore, infusion
sets of different brand and model may
differ in the detected air detection.
Note 2: The above declared air detection is
based on BOON A2 and B.Braun Intrafix
Safeset infusion sets, temperature of
℃
Accumulate air 0.1-4ml 1.5ml
KVO Rate 0.1-5.0ml/h 0.5ml/h
Time Near End
Off, 1-30 min
when the time is <10min, step for 1min,
and step for 5min when the time is ≥10min.
3min
Max. Rate Limit 0.1-2000ml/h 2000ml/h
Drip Setting
Drip:10-60drop/ml 20drop/ml
gtt/min: On, Off Off
Drip Abnormal On, Off Off
Empty
High, Low Low
Auto-lock Time Off, 1-5min, step for 1min Off
Reminder Time Off, 1-5min, step for 1min 2min
Weight Unit kg, lb kg
Sound Volume 1-8 4
Brightness 1-8 4
Night Mode On, Off Off
Nurse call On, Off Off
History Record Can store up to 2000 records. /
Brand Selection On, Off On
BW Mode
Configuration
Conc., Drug Amount and Volume Conc.
Drug Library On, Off Off
Para. Memory On, Off Off
Date and Time
Time:_ _:_ _ 00:00
Date:_ _ _ _-_ _-_ _ 01-01-2018
Time format: 12h, 24h 24h