A-7
Date format: yyyy-mm-dd, mm-dd-yyyy or
dd-mm-yyyy
Domestic: yyyy-mm-dd
International:
dd-mm-yyyy
Language
You can select language according to
actual needs
/
Pressure Unit mmHg, kPa, bar and psi mmHg
Alarm Sound Sound1, Sound2, Sound3 Sound2
Anti-bolus
switch
On, Off On
Drop sensor On, Off Off
Infusion
accuracy
Infusion pump with conventional infusion
sets: Infusion accuracy of the whole rate
≤±5%
Note 1: The infusion accuracy of the
infusion system (infusion pump and
infusion set) is affected by inner and outer
diameter, material, elasticity of the infusion
set and other factors. Therefore, infusion
sets of different brand and model may
differ in the infusion accuracy.
Note 2: The above declared infusion
accuracy is based on BOON A2 and
B.Braun Intrafix Safeset infusion sets,
temperature of 20±2.
Infusion pump with silicone specified
infusion sets: Infusion accuracy of the
whole rate ≤±5%, temperature of 20±2.
Infusion pump with PVC specified infusion
sets: Infusion has run for 24h or has
pumped 3L of fluid, infusion accuracy
≤±10%, temperature of 20±2
/
Dose of single
fault
0.25ml
Note: The above declared single fault is
based on BOON A2 infusion set at a rate of
25ml/h
/
Status
indicators
Stop, running, bolus, KVO, pause,
standby, alarm and purge
/
Alarm
Information
See complete information in
C Alarm
Information
/