4-3
Load infusion set according to the following method:
1. Pull the door latch up to open
the door.
2. Push up the free flow clamp.
3. Load the set, confirm it is
firmly loaded into the tube slot,
and adjust or press the set at
the gray zone, to confirm the set
is loaded correctly.
4. Push the door latch and close
the door gently, the system will
enter the [Set Selection]
interface, indicating that the
infusion set is loaded correctly;
otherwise, it needs to be
reloaded.
WARNING
The infusion set should be firmly loaded into the slot, and not jutting on the
outside of the slot.
Before using this infusion pump, the infusion pump, infusion set and other
accessories should be loaded correctly.