4-6                            BeneVision N22/BeneVision N19 Patient Monitor Service Manual 
 
4.7.4 Alarm Failures 
Fault Symptom  Possible Cause  Countermeasure 
Alarm LED off or cannot be 
turned off while the audible 
alarm is sounding 
Cable defective or improperly 
connected 
1. Verify the cable connecting the alarm LED board to 
the front housing interface board, and the cable 
connecting the front housing interface board to the 
main board are properly connected.   
2. Verify the cables and connectors are not damaged. 
Alarm LED board failure  Replace the alarm LED board. 
Front housing interface board failure  Replace the front housing interface board. 
Main board failure  Replace the main board. 
No audible alarm sounds 
emitted while the alarm LED 
is normal 
Audible alarm disabled 
Select the Main Menu quick key, from the System 
column, select Maintenance  →  enter required 
password  →  select Alarm, set Minimum Alarm 
Volume to a proper value. Select the Main Menu 
quick key, from the Alarm column, select Setup, 
adjust the alarm volume to a proper value. 
Cable defective or improperly 
connected 
1. Verify the cable connecting the speaker to the 
main board is properly connected. 
2. Verify the cables and connectors are not damaged. 
Speaker failure  Replace the speaker. 
Main board failure  Replace the main board.