8
Operating the System
CAUTION
Operate the system strictly as instructed by this manual.
Inappropriate use of the system may lead to unreliable test results or
even equipment damage or personal injury.
Before using the system for the first time, run the calibration program
and QC program to make sure the system is in normal status.
Be sure to run the QC program every time you use the system,
otherwise the result may be unreliable.
Do not open the covers of the sample/reagent disk cover when the
system is in operation.
Do not open the reaction disk cover when system is in operation.
The RS-232 port on the analyzing unit is to be used for connection
with the operation unit only. Do not use it for other connections. Only
use the supplied cable for the connection.
The operation unit is a personal computer with the
BS-120/BS-130/BS-180/BS-190 operating software installed.
Installing other software or hardware on this computer may interfere
with the system operation. Do not run other software when the
system is working.
Computer virus may destroy the operating software or test data. Do
not use this computer for other purposes or connect it to the Internet.
Do not touch the display, mouse or keyboard with wet hands or
hands with chemicals.
Do not place the Main Power to ON again within 10 seconds since
placing it to OFF; otherwise the system may enter protection status.
If it does so, switch off the Main Power and switch it on again.
Service and Maintenance
CAUTION
Maintain the system strictly as instructed by this manual.
Inappropriate maintenance may lead to unreliable results, or even
equipment damage and personal injury.
Dust may accumulate on the system surface when the system is
exposed to the outside for a long time.To wipe off dust from the
system surface, use a soft, clean and wet (not too wet) cloth, soaked
with mild soap solution if necessary, to clean the surface. Do not use
such organic solvents as ethanol for cleaning. After cleaning, wipe
the surface with dry cloth.
Switch off all the powers and unplug the power cord before cleaning.
Take necessary measures to prevent water ingression into the
system, otherwise it may lead to equipment damage or personal
injury.
Replacement of such major parts as lamp, photometer, sample
probe, mixer and syringe plunger assembly must be followed by a
calibration.

Table of Contents

Other manuals for Mindray BS-120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BS-120 and is the answer not in the manual?

Mindray BS-120 Specifications

General IconGeneral
BrandMindray
ModelBS-120
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals