9 Troubleshooting
9-4
Severity: Invalidating module (Reserved)
Level
Description Actions taken by the system
12 Errors to invalidate
ISE tests
The ISE tests that are affected by bubbled sample or go out of measurement range due to
abnormal(insufficiently-aspirated) sample are invalidated and then run aga
in. If the error occurs
continuously for 3 times due to the same cause, the system will skip all ISE tests related to the
sample during the current batch of tests.
Severity: Forbidding module (Reserved)
Level
Description Actions taken by the system
13 Errors to forbid
sample bar code
scanning
When the sample bar code reader goes wrong and cannot scan sample bar code label normally,
the system will not try again during the measurement. The sample positions should be set up
manually. Only when reconnected and proved to be normal, the sample bar code reader can start
working.
14 Errors to forbid ISE
When errors (like calibration slope out of range) occur due to failed ISE component, ISE tests will
be forbidden. If such errors occur during measurement, the system will invalidate the failed ISE
tests and skip all other ISE tests in current batch. If such errors occur in idle status, the system
will not include ISE tests once starting analysis.
Severity: Errors to forbid LIS (Reserved)
Level
Description Actions taken by the system
15 Errors to forbid LIS When the LIS host goes wrong or the network connection and settings are improper, the
system cannot download sample information from or send test results to LIS. You can use all
functions associated to LIS only after reconnecting the system to LIS successfully.