9 Troubleshooting 9-13
100640040007
Main unit result error:
Shaking hands with
other units error
Something wrong with the
connection cable or
controlling unit is
executing other operations
Check whether the mainboard connection is OK. Restart the
analyzing unit and reopen the operating software. If it
repeats for several times, please contact the R&D engineer.
100640050007
Main unit result error:
E2PROM read error
E2PROM read error
Restart the analyzing unit and the operating software.
Execute failure recovery. If the error remains, replace the
mainboard.
100640080007
Main unit result error:
E2PROM write error
E2PROM write error
Restart the analyzing unit and the operating software.
Execute failure recovery. If the error remains, replace the
mainboard.
100640090007
Main unit result error:
Downloading parameter
Main unit downloading
parameters to subunits;
unable to response
Execute the previous operation after waiting for 30-60
seconds. If there is no response for a long time, execute
mechanical resetting or check the connection of the
mainboard.
100640100007
Main unit result error:
Parameter downloading
failed
Mainboard connection
error or wrong parameter
downloading or wrong
parameter configuration
Check the connection of the mainboard.Reopen the
operating software after restarting the analyzing unit.If the
error remains, replace the mainboard.
100640110007
Main unit result error:
Resetting other units
Resetting other units;
unable to response.
Execute the previous operation after waiting for 30-60
seconds. If there is no response for a long time, execute
mechanical resetting or check the connection of the
mainboard.
100640120007
Main unit result error:
Other units resetting
error
Other units resetting error
Execute failure recovery. If ithe error repeats continuously
for more than 3 times, please contact the R&D engineer.
100640150007
Main unit result error:
Invalid status.
Self-check
Self-checking; Other
operations not supported
Run the previous operation after waiting for 30-60
seconds.If there is no response for a long time, execute
mechanical resetting or restart analyzing unit. If ithe error
repeats for several times, please contact the R&D engineer.
100640150017
Main unit result error:
Invalid status. Error
Main unit error; other
operations not supported
Execute failure recovery; if the error repeats, replace the
mainboard.
100640160007
Main unit result error:
Unit busy. No response
Busy. No response
Run the previous operation after waiting for 30-60 seconds.
If there is no response for a long time, execute mechanical
resetting or restart analyzing unit and the operating
software.