9 Troubleshooting 9-15
100680040015
Sample unit result error:
Status error. Error
sample unit error; current
operations not supported
Execute failure recovery. If not successful, check the
sample probe, sample/reagent disk diriver motor/sensor
and the connection and then execute failure recovery.
100680050005
Sample unit result error:
Unit busy. No response
Unit busy. No response
Execute the previous operation after waiting for 30-60
seconds. If there is no response for a long time, restart the
analyzing unit and reopen the operating software.
100680090005
Sample unit result error:
Wrong configuration
parameter
Parameter configuration
error
Configure the right parameter.
100680170005
Sample unit result error:
Fluid controlling error
Fluid path control error
Check the connection of DI water and waste sensor. Check
whether the syringe is leaking, or there is hanging liquid on
the sample probe, the pump-valve assembly is wrong.
Restart the analyzing unit and open the operating software.
Adjust the fluid path function by using Alignment.
100680180005
Sample unit result error:
Sample disk rotation
error. Cannot reach the
home position
Sample disk rotation error
Check the connection of sample/reagent disk motor/sensor.
Execute failure recovery.
100680180015
Sample unit result error:
Sample disk rotation
error. Cannot move
away from the home
position
Sample disk rotation error
Check the connection of sample/reagent disk motor/sensor.
Execute failure recovery.
100680180025
Sample unit result error:
Sample disk rotation
error. Step missing
Sample disk rotation error
Check the sample/reagent disk diriver motor/sensor and the
connection. Check whether the rotation of the
sample/reagent disk is blocked. Execute failure recovery.
100680180035
Sample unit result error:
Sample disk rotation
error. Probe in disk
Sample disk rotation error.
Probe in the disk
Execute sample disk rotation after moving the sampole
probe out of the sample disk by using Alignment.