9 Troubleshooting                                                                                    9-17 
100680210005 
Sample unit result error: 
Sample probe horizontal 
movement error. Cannot 
reach the home position
Sample  probe  horizontal 
movement  error.  Cannot 
reach the home position 
Check the connection between probe horizontal movement 
motor  and  the  sensor.  Check  whether  the  sample  probe 
horizontal  movement  is  blocked.  Execute  mechanical 
resetting. 
100680210015 
Sample unit result error: 
Sample probe horizontal 
movement error. Cannot 
move  away  from  the 
home position 
Sample  probe  horizontal 
movement  error.  Cannot 
leave the home position 
Check the connection between probe horizontal movement 
motor  and  the  sensor.  Check  whether  the  sample  probe 
horizontal  movement  is  blocked.  Execute  mechanical 
resetting. 
100680210055 
Sample unit result error: 
Sample probe horizontal 
movement  error.  Step 
missing 
Sample  probe  horizontal 
movement  error.  Step 
missing. 
Check  the  movement  of  the  sample  probe  by  using 
Alignment. Check whether the horizontal movement of the 
sample probe is blocked. Execute mechanical resetting. 
100680210065 
Sample unit result error: 
Sample probe horizontal 
movement  error. 
Disabled 
Sample  probe  horizontal 
movement  error. 
Forbidden. 
Execute mechanical resetting. 
100680220005 
Sample unit result error: 
Sample  probe  vertical 
movement error. Cannot 
reach the home position
Sample  probe  vertical 
movement  error.  Cannot 
reach the home position 
Check  the  connection  between  probe  vertical  movement 
motor  and  the  sensor.  Check  whether  the  sample  probe 
vertical  movement  is  blocked.  Execute  mechanical 
resetting. 
100680220015 
Sample unit result error: 
Sample  probe  vertical 
movement error. Cannot 
move  away  from  the 
home position 
Sample  probe  vertical 
movement  error.  Cannot 
leave the home position 
Check  the  connection  between  probe  vertical  movement 
motor  and  the  sensor.  Check  whether  the  sample  probe 
vertical  movement  is  blocked.  Execute  mechanical 
resetting. 
100680220025 
Sample unit result error: 
Sample  probe  vertical 
movement  error.  Trying 
to  move  away  from  the 
initial limit position 
Sample  probe  vertical 
movement error. Trying to 
move away from the initial 
limit position 
Execute mechanical resetting. If  the error remains, check 
the status of the sensor.