9 Troubleshooting   
9-24 
100660080000 
Temperature  unit result 
error:  Undefined 
temperature parameter 
Wrong  system  operation 
parameter 
Configure the right parameter. 
100660090000 
Temperature  unit result 
error:  Wrong 
temperature parameter 
Wrong  system  operation 
parameter 
Configure the right parameter. 
100660100000 
Temperature  unit result 
error: Undefined sensor 
parameter 
Undefined  sensor 
parameter 
Configure the right parameter. 
100660110000 
Temperature  unit result 
error:  Wrong  sensor 
parameter 
Wrong sensor parameter  Configure the right parameter. 
100660120000 
Temperature  unit result 
error:  Undefined  target 
temperature 
Undefined  target 
temperature 
Configure the right parameter. 
100660130000 
Temperature  unit result 
error:  Wrong  target 
temperature parameter 
Wrong  target  temperature 
parameter 
Configure the right parameter. 
100660160000 
Temperature  unit result 
error:  Parameter  write 
protection 
Unit  parameter  write 
protection.  Parameter 
configuration forbidden   
Configure the  parameter after parameter configuration is 
permitted. 
100661250000 
Temperature  unit result 
error: Main unit sending 
error 
/ 
Check the connection of the temperature unit. Restart the 
operating  software  and  retry  the  command.  If  the  error 
repeats for several times, check the cable and the board. 
100661270000 
Temperature  unit result 
error:  Main  unit  does 
not  receive  the 
Temperature unit result 
/ 
Check the connection of the temperature unit. Restart the 
operating  software  and  retry  the  command.  If  the  error 
repeats for several times, check the cable and the board.