9 Troubleshooting 9-27
100670170095
Mixing unit result error:
Mixing bar vertical
movement error.
Reaction disk is rotating
Reaction disk rotating.
Mixing not permitted
Run mechanical resetting or mixing unit mechanical
resetting. Rotate the reaction disk after mixing bar leaves
the cuvettes.
100670200005
Mixing unit result error:
Write protection
Unit parameter write
protection. Parameter
configuration forbidden
Configure parameter after iparameter configuration is
permitted.
100671250005
Mixing unit result error:
Main unit sending error
/
Check the connection of the mixing unit. Restart the
operating software and retry the command. If the error
repeats for several times, check the cable and the board.
100671270005
Mixing unit result error:
Main unit does not
receive the mixing unit
result
/
Check the connection of the mixing unit. Restart the
operating software and retry the command. If the error
repeats for several times, check the cable and the board.