9 Troubleshooting 9-31
10070001CLM5
IISE unit result error:
Clean cycle error
No ISE wash solution Refill ISE wash solution
10070001COM5
ISE unit result error:
Communication error
ISE connection error Reconnect; reset ISE; turn on the power of ISE module .
10070001GAA5
ISE unit result error:
Purge A cycle error. Air
in calibrant A
Calibrator A bottle empty;
tube not connected;
pump A not working;
tube blocked, broken or
twisted;
fibrin and salt in the tube;
air bubble detector error;
waste pump error
Wrong electrode installation. Check the pressing plate
spring and the sealing to ensure all the electrodes and
the O-ring are in the right place.
Wash the module by using <CLEN > program. Remove
the module and then wash it. Reinstall the sensor.
Replace air bubble detection sensor;
replace the waste pump;
replace the calibrator A and recalibrate;
reconnect or replace the tube;
check the connection of the motor;
replace the pump box; replace the motor;
replace the tube.
10070001GAF5
ISE unit result error:
Purge A cycle error. No
flow
Calibrator A bottle empty;
wrong electrode installation
Replace the calibrator A, then recalibrate; check the
installation of the electrodes.