9 Troubleshooting 9-39
1007000300A5
ISE unit result error: Cl,
Na electrodes voltage
overflow (Cal A in calib
mode, Cal B in urine
mode)
Electrodes performance
degrading.
Replace the electrodes and or
calibrator A
Run test after replacing problematic electrodes;
Check the calibrator A and recalibrate; there might be
drift, if it is a new electrode. Rerun after 15 minutes.
1007000300C5
ISE unit result error: Cl,
K electrodes voltage
overflow (Cal A in calib
mode, Cal B in urine
mode)
Electrodes performance
degrading.
Replace the electrodes and or
calibrator A
Run test after replacing problematic electrodes;
Check the calibrator A and recalibrate; there might be
drift, if it is a new electrode. Rerun after 15 minutes.
1007000300E5
ISE unit result error: Cl,
K, Na electrodes voltage
overflow (Cal A in calib
mode, Cal B in urine
mode)
Electrodes performance
degrading.
Replace the electrodes and or
calibrator A
Run test after replacing problematic electrodes;
Check the calibrator A and recalibrate; there might be
drift, if it is a new electrode. Rerun after 15 minutes.
100700040025
ISE unit result error: Na
electrode voltage noise
(Cal B/Sample)
Electrodes performance
degrading.
Replace the electrodes and or
calibrator A
Run test after replacing problematic electrodes;
Check the calibrator A and recalibrate; there might be
drift, if it is a new electrode. Rerun after 15 minutes.
100700040045
ISE unit result error: K
electrode voltage noise
(Cal B/Sample)
Electrodes performance
degrading.
Replace the electrodes and or
calibrator A
Run test after replacing problematic electrodes;
Check the calibrator A and recalibrate; there might be
drift, if it is a new electrode. Rerun after 15 minutes.
100700040065
ISE unit result error: K,
Na electrodes voltage
noise (Cal B/Sample)
Electrodes performance
degrading.
Replace the electrodes and or
calibrator A
Run test after replacing problematic electrodes;
Check the calibrator A and recalibrate; there might be
drift, if it is a new electrode. Rerun after 15 minutes.