EasyManua.ls Logo

Mindray BS-120 - Page 250

Mindray BS-120
267 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 Calculation Methods 10-7
21
1
VSV
SV
++
+
- double-reagent volume calibration factor
Double-wavelength (for both single-reagent and double-reagent)
The calculation method is similar to that for single-wavelength reaction, except that in
every measurement period the absorbance is the difference between primary wavelength
absorbance and secondary wavelength absorbance.
10.2.2.2 Calculating Response of Fixed-time Reaction
Single-wavelength (for both single-reagent and double-reagent)
bs
RRR
=
s
R
and
b
R
are calculated through
km
tt
tt
AA
R
km
=
.
Where,
s
R
- original response
b
R
- reagent blank response.
b
R
will be the response of the latest reagent blank. If no
reagent blank has been required,
b
R
0
l
t
- start time for absorbance reading
m
t
- end time for absorbance reading
Double-wavelength (for both single-reagent and double-reagent)
The calculation method is similar to that for single-wavelength reaction, except for every
measurement period, the absorbance is the difference between primary wavelength
absorbance and secondary wavelength absorbance.
10.2.2.3 Calculating Response of Kinetic Reaction
Single-wavelength (for both single-reagent and double-reagent)
bs
RRR
=
s
R
and
b
R
are calculated through the method of least squares.
Where,
s
R
- original response
b
R
- reagent blank response.
b
R
will be the response of the latest reagent blank. If no
reagent blank has been required,
b
R
0
Formula with the method of least squares:

Table of Contents