3 Installation 3-16
normally, that is, to follow the exiting prompt shown in dialog box to complete
the exiting procedures, including cuvettes replacement, washing, turning off
the light and so on.
Figure 3-10 Emergent Exit Dialog Box
After exiting the operating software, turn off the power of the printer, the
operation unit (personal computer), the display and the analyzing unit
(analyzer) in turn.
The main power of analyzer should be kept on if the reagents are held in the
sample/reagent disk so as to keep cooling the reagents. If the reagents are
taken out to store in other place, the main power could be turned off.
NOTE
The refrigerator still functions after the Power is turned off. To shut
down the refrigerator, turn off the Main Power.