EasyManua.ls Logo

Mindray BS-120 - Page 96

Mindray BS-120
267 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 Units Description
4-41
4.8.3.1 ISE Module
Figure 4-46 ISE Module
The dismounting/mounting steps are as follows (see Figure 4-46 ).
Dismounting steps:
1. Remove the table panel and the ISE unit door.
2. Unplug the draining tube connected to the ISE module.
3. Loosen the screw and remove the ISE shielding cover.
4. Loosen the four M3X8 cross pan head screws retained on the ISE shield
housing and remove the ISE module.
Mounting steps:
1. Install the ISE module on the ISE shield housing by tightening the four M3X8
cross pan head screws.
2. Install the ISE shielding cover by tightening the screw.
3. Pull out the draining tube through the hole in the ISE shielding cover.(it is not
indicated in figure)
4. Install the table panel and ISE unit door.