Page #124 background imageLoading...
Page #124 background image
5-44 Image Optimization
Parameter Description
Opacity
Function: to adjust the transparency value for 3D image rendering. It implies the
transparency of the light. The higher the value is, the tougher the surface
becomes.
Range: 0% to 100%.
The lower the number is, the more transparent the gray scale information will be.
Available only in Surface rendering mode.
Smooth
Function: to set the smoothness of 3D image.
Range: 0 to 10. 0 refers to no smooth effect; 1-10 provides 10 effects in
incremental order.
Tip: insufficient smoothness can result in a fuzzy image; however, too much
smoothness will lead to image distortion.
Brightness
Function: to set the brightness of image.
Range: 0%-100%, in increments of 2%.0% represents the minimum brightness,
while 100% represents the maximum.
NOTE: the adjustment for 3D (VR) and MPR.
Contrast
Function: set the contrast scale of the image (contrast).
As long as the contrast becomes larger, the bright spot and dark spot on the
image change as well.
Range: 0%-100%, in increments of 2%.
NOTE: the adjustment for 3D (VR) and MPR
Tint
Function: enable/disable tint map. The color of image change according the tint
value.
Selection: OFF, 1-8.
Gray Map
Function: enable/disable gray map. The color of image change according the
gray scale value.
Adjusting level: tap [MPR]. Rotate [Gray Map] knob. The adjusting range: 1-8.
VR Orientation
To quickly rotate the VR.
Selection: 0°, 90°, 180°, 270°.
3D iClear
Function: The 3D adaptive filtering is used to enhance the spatial continuity of
tissue structure, and restrain speckle noise to increase image contrast, so as to
optimize volume image data.
Selection: OFF, 1-7.
Select VR imgae. Adjust 3D iClear to optimize VR image.
Select MPR imgae. Adjust 3D iClear to optimize MPR image.
Face+
Optimize the signal of the face area to reduce the noise of the AF area and
improve the signal-noise ratio of the face area, make the face more fullness to
easily obtain the image of fetal face.
Range: 1-3.
VR Refine
Function: Optimize the signal-noise ratio and the contrast of VR image.
Selection: OFF, 1-7.
MagiClean
Function: Reduce the noise and keep the boundary of the image more clearly.
Selection: only enabled in rendering mode Off, Low, Mid, High and Max.

Table of Contents

Other manuals for Mindray DC-80A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-80A and is the answer not in the manual?

Mindray DC-80A Specifications

General IconGeneral
BrandMindray
ModelDC-80A
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals