Page #247 background imageLoading...
Page #247 background image
DICOM/HL7 11-5
DICOM storage preset items are described as follows:
Name Description
Configure
New
Service
Device
After setting the servers in the DICOM Preset screen, the
names will appear in the drop-down list. Select the name of
the storage server.
Service Name The default is xxx-Storage, user-changeable.
AE Title
Application Entity title. It should be consistent with that of the
storage server.
Port
DICOM communication port, 104 is the default. The port
should be consistent with that of the storage server port.
Maximum Retries
Set the maximum retries (0-9). The default value is 3. If the
DICOM task sending to the server fails, the retry times
should be 3.
Interval Time(Sec) Interval time for the system to retry a connection.
Timeout(Sec)
Refers to the amount of time after which the system will stop
trying to establish a connection to the service.
TLS
Transport Layer Security. Select whether to encrypt the data
during network transportation.
Cine Zoom Mode Select the cine zoom mode during image file storage.
Compression Mode
Select the compression mode: uncompressed, RLE (the
image not compressed), JPEG and JPEG2000.
Compression Ratio
Select the JPEG compression ratio: lossless, low, medium
and high.

Table of Contents

Other manuals for Mindray DC-80A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-80A and is the answer not in the manual?

Mindray DC-80A Specifications

General IconGeneral
BrandMindray
ModelDC-80A
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals