7-4 Physiological Signal
7.2 Respiratory Wave
7.2.1 Respiratory Wave Operation Basic Procedures
1. Connect the ECG lead and position the ECG electrodes. For details of ECG lead connection,
see ā€œ7.1.1 ECG Operation Basic Proceduresā€.
2. Press <F12 Physio> on the control panel to enter Physio screen.
3. Switch the imaging modes and display formats, adjusting the parameters to get an optimized
image.
4. Parameter adjusting
In Physio menu, click [RESP] on the touch screen. Adjust [Speed], [RESP Gain], [Position] and
[Invert].
5. Exit Respiratory display mode, and remove ECG electrodes from the patient.
6. Press <F12 Physio> to exit physio mode.
7.2.2 Parameter Description
The physio parameters are described as follows:
Parameter
Description
ECG Source Select ECG source.
Gain
Function: to set the amplitude of the trace.
Method: click [Gain] on the soft menu, or press/rotate the corresponding knob.
Each adjustment takes you to the next setting.
Value: 0-30, in increments of 1.
Position
Function: to set the vertical position of the traces on the image display.
Method: select [Position] on the touch screen. Each adjustment cycles you to the
next setting.
Value: 0%-100%, in increments of 5%.
Speed
Function: change the speed of the physio trace.
Value: 20-145mm/s.
T1 Function: to set the delay time T1 in Single trigger or Dual trigger.
T2
Function: to set the delay time T2.
Method: select a proper value in the drop-down list.
Interval
Function: to set the time interval for Timer.
Method: select a proper value.
Invert To invert the display.