mindray
DC7 Diagnostic Ultrasound System
9
You can click [Delete] to delete the item in Service List.
1.4.2 DICOM Print Preset
Figure 5 DICOM Storage Preset Screen
The main parameters of DICOM Print are described as below:
Name Description
Devices
After you set the server (s) in DICOM Server screen, the name (s)
will appear in the drop-down list, select the name of the print
server.
Service name Default is xxx-Print, user-changeable.
AE Title
Application Entity title, here, it should be consistent with that of the
print server.
Port
DICOM communication port, 104 is default. Here, the port should
be consistent with that of the print server port.
Maximum
Retries
09 times are optional, the default is 3. if the DICOM storage task
that is sent to this server is failed, the task will be retried
automatically, the retry times is set here.
Interval Time
Sec
51800 seconds, 15 seconds is default. (Reserved)
TimeoutSec
Refers to timeout during association establishment.
Value: 5-60s, in increments of 5s, and 15 is default.
Copies
Refer to copies of printed files. You can select among 1 through 5,

Table of Contents