(3) Film Size: 8X10
(4) MAGNIFICATION TYPE: NONE/CUBIC/ REPLICATE/BILINEAR
(5) AE Title: Print_Scp (recommended)
2. Method 2
(1) Settings: MONOCHROME2
(2) Medium Type: Blue Film
(3) Film Size: 14X17
(4) MAGNIFICATION TYPE: NONE/CUBIC/ REPLICATE/BILINEAR
(5) AE Title: Print_Scp (recommended)
The following are the incorrect settings of the ultrasound system DICOM print
preset:
Color printing:
1)
Settings:
MONOCHROME2
Incorrect.
“RGB” should be selected in color printing while
“MONOCHROME2” for B/W printing
2)
Medium Type: Blue Film
Incorrect.
Paper and Clear file are supported by color printer
according to the printer conformance statement; but
only Paper is available as displayed in the printer
screen.
3)
Film Size: A3
Incorrect.
Only three film size are available according to printer
screen display:
A4
14X17
8X10
4)
MAGNIFICATION TYPE
Can be set as: NONE/ CUBIC/ REPLICATE/
BILINEAR
5) AE Title: 14x17-dvfb
Incorrect.
Setting AE Title as “14x17-dvfb” will enforce the 14x17
DirectVista film (blue) printing.
3.12.2 Example 2
This example is being performed with FUJIFilm drypix4000 printer connected with the
ultrasound system.
DICOM print settings on ultrasound system:
1) Settings: MONOCHROME2