DC7 Diagnostic Ultrasound System
34
3.2 Common Failure: Server disconnects
the ultrasound system during
implementing storage
Potential Reasons:
The JPG compression is enabled on local device but is not supported on the
server;
Wrong AE_TITLE settings on local device and the server.
Troubleshooting:
Select “uncompressed” in DICOM Service -> Storage;
Set the correct local AE_TITLE and server AE_TITLE on local device according
to the DICOM Conformance Statement or settings of the server.
3.3 Common Failure: Network
disconnected (shown by network
icon) during transmission
Potential Reasons:
The network cable is disconnected during transmission;
The connected switch, router or hub is down or in failure;
Troubleshooting:
Connect the network cable;
Reset the switch, router or hub; or replace a network device with higher
performance.
3.4 Common failure: Worklist query
succeeded, but no patient is shown
Potential Reasons:
Scheduled Station AE Title in the preset is not consistent with the setting on
server;
Records on the server database are not accordant with the query conditions on
local ultrasound system
Troubleshooting: