mindray
DC7 Diagnostic Ultrasound System
14
Verify
Click to verify if the two DICOM application entities are normally
connected.
Default
Select an item in the service list, click [Default] and you can see “Y” in
the Default column.
Exit Click to exit the preset screen.
Tip: You should set MPPS as default server when you use it.
1.4.5 Storage Commitment Preset
Figure 8 Storage Commitment Preset Screen
The main parameters of storage commitment are described as below:
Name Description
Devices
After you set the server (s) in DICOM Server Preset screen, the
name (s) will appear in the drop-down list, select the name of the
server.
Service name Default is xxx-SC, user-changeable.
AE Title
Application Entity title, here, it should be consistent with that of
the storage commitment server.
Port
DICOM communication port, 104 is default. Here, the port should
be consistent with that of the storage commitment server port.
Maximum Retries
09 times are optional, the default is 3. (Reserved)
Interval Time
51800 seconds, 15 seconds is default. (Reserved)