7-4 Measurement
7.4 Measurement Accuracy
Table 1 Error of 2D Images
Within ±3%; or when the measured value is less
than 40 mm, the error is less than1.5 mm.
Within ±7%; or when the measured value is less
than 16 cm
2
, the error is less than 1.2 cm
2
.
Within ±7%; or when the measured value is less
than 16 cm
2
, the error is less than 1.2 cm
2
.
Micro-convex probe: within ±1 5%
Other probe: within ±10%
Table 2 Volume Measurements
Within ±10 %; or when the measured value is less than 64 cm
3
,
the error is less than 6.4 cm
3
.
Table 3 Time/Motion Measurements
Within ±3 %; or when the measured value is less
than 40mm, the error is less than 1.5mm.
When the correction angle is ≤60º, the PW
velocity ≤±10%.
Within the selected field range, the measurement accuracy is ensured within the
range mentioned above. The accuracy specifications are performance in the worst
conditions, or based on the real test for the system, regardless of acoustic speed
error.