Image Optimization 5-3
In real-time scanning of all image modes, press <B> on the control panel to return to B mode.
5.4.2 B Mode Parameters
In B mode scanning, the image parameter area in the upper left corner of the screen
displays the real-time parameter values as follows:
5.4.3 B Mode Image Optimization
Gain
To adjust the gain of the whole receiving information in B mode. The real-time
gain value is displayed in the image parameter area in the upper left corner of
the screen.
Rotate the <Gain/iTouch> knob clockwise to increase the gain, and
anticlockwise to decrease.
The adjusting range is 0-100.
Increasing the gain will brighten the image and you can see more received
signals. However, noise may also be increased.
Depth
This function is used to adjust the display depth of sampling, the real-time
value of which is displayed on the image parameter area in the upper left
corner of the screen.
Use the <Depth/Zoom> knob to adjust the depth.
The adjustable depth values vary depending upon the probe types.
Increase the depth to see tissue in deeper locations, while decrease the depth
to see tissue in shallower locations.
Depth increase will cause a decrease in the frame rate.