12-12 Probes and Biopsy
Before safety and performance is affected, probe 75LT38EA can be sterilized by Cidex
Activated Glutaraldehyde Solution for at least 217 times (10 hours for one time).
Refer to the instructions provided by the chemical manufacturer concerning
concentration of the sterilization solution, method of sterilization and dilution and
cautions during use.
Do not soak the probe connector or the cable near it into water or any solution.
Follow local regulations when selecting and using the sterilization solution.
3. Rinse the probe with plenty of sterile water (about 2 gallons) for at least 1 minute to
remove all chemical residues on it. Or, follow the rinsing method recommended by the
sterilization solution manufacturer to rinse the probe.
4. Wipe off the water on the probe with sterile cloth or gauze after rinsing it. Do not dry the
probe by heating.
Immerse the probe 75LT38EA in the solution
Repeated disinfection may degrade the performance and safety of the probe.