Probes and Biopsy 12-11
NOTE:
1.
Observe the graph here carefully to immerse the transducer. Only soak parts
of the transducer below the strain relief.
2.
Repeated disinfection will eventually damage the probe, please check the
probe performance periodically.
Compatible Disinfectants
Please see Mindray Transducer Disinfectant Recommendation for the disinfectant use in
details.
Sterilization
For intra-operative probes (75LT38EA), they have to be sterilized after completing each
examination.
1. Wear sterile gloves to prevent infection.
2. Clean the probe before sterilizing it. MINDRAY recommends the following solutions to
sterilize the probe.
Hydrogen Peroxide and Peroxyacetic Acid -based sterilization solution
Trade Name
Chemical Name
Procedures
Minncare
®
Cold
Sterilant
22% Hydrogen
Peroxide
4.5% Peroxyacetic Acid
Dilute the sterilant with sterilized purified
water (1:20).
Immersed time: 11 hours.
Temperature: 20-25.
Please refer to the instructions provided by
the manufacturer of the solution for details.
Before safety and performance is affected, probe 75LT38EA can be sterilized by Minncare
COLD STERILANT for at least 135 times (11 hours for one time).
Glutaraldehyde-based sterilization solution
Trade Name
Chemical Name
Procedures
Cidex Activated
Glutaraldehyde
Solution
2.2-2.7%
Glutaraldehyde
Soak the transducer into the activated
solution for 10 hours (20-25)
Please refer to the instructions provided by
the manufacturer of the solution for details.
Strain relief
Probe handle