10-2 DICOM
1. Press <Setup> to enter the [Setup] menu.
2. Select [Network Preset].
3. Local TCP/ IP preset items are described as follows:
Name
Description
Current Net Adapter
To select network connection mode.
DHCP
/ Static
DHCP: IP address will be automatically obtained from DNS server;
Static: you need to enter the IP address.
IP Address
IP address of the system.
Subnet Mask
Used to set different network segment.
Gateway
Used to set the gateway IP.
Note:
The IP address of the system should not be occupied by other devices in the network,
or the DICOM function will fail.
After you changed “Station Name”, you need to restart the system to make the setting
effective (press the power button and select “Shut Down” and restart the system after
a while).
10.1.2 DICOM Preset
To set the DICOM service and server properties.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DP-50 Vet and is the answer not in the manual?

Mindray DP-50 Vet Specifications

General IconGeneral
Transducer Ports2
Cine LoopYes
Operating Temperature10°C to 40°C
Image StorageUSB
ApplicationsCardiology
Probe TypesLinear, Convex, Micro-convex
Humidity Range30% to 75% (non-condensing)