Probes and Biopsy 12-31
NGB-016
1. Put on the sterile transducer sheath.
2. Select a proper needle-guided bracket, and match the groove with the tab of the
transducer. Mount the bracket onto the transducer.
3. Screw the pinch nut of the needle-guided bracket to confirm that the needle-guided
bracket is properly installed on the transducer.
4. Select a proper guiding block and push it into the groove above the angle block, and
clamp it tightly.
5. Screw the nut of the block to secure the block.
6. Insert a biopsy needle with the same specification as that of the guiding block into the
hole of the guiding block.
CAUTION:
Ensure that all guide parts are seated properly prior to performing
a biopsy.
12.2.5 Biopsy Menu
Press <Biopsy> to show the biopsy menu.