Probes and Biopsy 12-27
NGB-004
1. Put on the sterile probe sheath.
2. Open the retaining clamp, align the needle-guided bracket with the transducer to locate
the locating bulge on the needle guide to the locating grooves on the transducer, and
then turn the retaining clamp to match it with the transducer.
3. When the retaining clamp is turned to the right position, the locking nut will lock the
retaining clamp and the needle-guided bracket is then mounted to the right position.
NGB-007
Metal
1) Put on the sterile transducer sheath.
2) Hold the transducer by one hand, select the proper needle-guided bracket, and
hold it with the other hand. Match the groove and tab with the tab and groove of
the transducer respectively. Amount the bracket onto the transducer.
3) Screw the pinch nut of the needle-guided bracket to confirm that the needle-
guided bracket is properly installed on the transducer.