Transducer Maximum Surface Temperature B-1
Appendix B Transducer
Maximum Surface
Temperature
According to the requirements of the section 201.11 in the standard IEC 60601-2-37: 2015,
the transducer surface temperature has been tested in two kinds of conditions: the transducer
suspended in still air or transducer contacting human-tissue mimicking material.
* The measurement data were obtained under the test conditions employed at Mindray.
Transducer model
Maximum surface temperature
C)
Contacting TMM
Maximum surface
temperature (°C)
Suspending in air
10L24EA
40.6
32.8
35C20EA
39.8
33.5
35C50EA
39.2
35.0
65C15EA
40.6
34.2
65EB10EA_T
41.4
30.3
65EB10EA_S
41.3
29.9
65EC10ED
41.1
29.5
65EL60EA
40.7
30.2
75L38EA
40.6
34.1
75L53EA
39.6
30.8
75LT38EA
39.8
28.3
50L60EAV
37.6
32.8
75L50EAV
40.4
31.6

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DP-50 Vet and is the answer not in the manual?

Mindray DP-50 Vet Specifications

General IconGeneral
Transducer Ports2
Cine LoopYes
Operating Temperature10°C to 40°C
Image StorageUSB
ApplicationsCardiology
Probe TypesLinear, Convex, Micro-convex
Humidity Range30% to 75% (non-condensing)

Related product manuals