C-2 Acoustic Output Reporting Table (60601-2-37)
C.1 10L24EA
Transducer Model: 10L24EA
Imaging Mode: M-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
1.07
0.20
0.40
0.31
Index component value
0.20
0.13
0.19
0.40
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
2.89
P (mW)
6.89
3.27
7.33
P
1
(mW)
6.89
3.27
z
s
(cm)
1.00
z
b
(cm)
1.08
z
MI
(cm)
1.08
z
pii
,α
(cm)
1.08
f
awf
(MHz)
7.29
6.03
5.87
6.07
Other
Informatio
n
prr (Hz)
1,000.0
0
srr (Hz)
/
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
265.84
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
44.69
I
spta
at z
pii
or z
sii
(mW/cm
2
)
80.30
p
r
at z
pii
(MPa)
3.65
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
12.9cm
12.9cm
12.9cm
12.9cm
Focus position
2.0cm
5.0cm
2.0cm
5.5cm
Working Frequency
10
H12.0
H12.0
H12.0
PRF
1000
2999
2999
2999