C-6 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 10L24EA
Imaging Mode: B+PW-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.99
0.19
1.04
0.78
Index component value
0.19
0.15
0.23
1.04
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
2.36
P (mW)
11.94
12.50
26.85
P
1
(mW)
6.03
8.07
z
s
(cm)
1.00
z
b
(cm)
0.79
z
MI
(cm)
0.79
z
pii
,α
(cm)
0.79
f
awf
(MHz)
5.75
7.05
7.11
7.15
Other
Informatio
n
prr (Hz)
699.00
srr (Hz)
58.00
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
251.55
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
111.04
I
spta
at z
pii
or z
sii
(mW/cm
2
)
157.37
p
r
at z
pii
(MPa)
2.67
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
6.5cm
12.9cm
12.9cm
12.9cm
B Focus Position
1.0cm
10.0cm
1.0cm
10.0cm
SV Position
1.0cm
10.0cm
1.0cm
10.0cm
B Working Frequency
8
10
8
8
PW Working Frequency
5.7
6.6
5.7
5.7
B PRF
7695
4199
1258
1258
PW PRF
699
699
3595
3595
PW SV
0.5mm
0.5mm
0.5mm
0.5mm