EasyManua.ls Logo

Mindray DP-50 Vet - Page 264

Mindray DP-50 Vet
343 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C-16 Acoustic Output Reporting Table (60601-2-37)
C.3 35C50EA
Transducer Model: 35C50EA
Imaging Mode: M-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
1.04
0.25
0.87
0.52
Index component value
0.19
0.25
0.47
0.87
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
1.64
P (mW)
30.81
27.84
41.00
P
1
(mW)
12.25
15.99
z
s
(cm)
2.89
z
b
(cm)
4.82
z
MI
(cm)
4.72
z
pii
,α
(cm)
4.72
f
awf
(MHz)
2.46
3.34
2.46
2.46
Other
Informatio
n
prr (Hz)
1,000.0
0
srr (Hz)
/
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
227.67
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
181.91
I
spta
at z
pii
or z
sii
(mW/cm
2
)
415.15
p
r
at z
pii
(MPa)
2.24
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
14.8cm
14.8cm
14.8cm
14.8cm
Focus position
6.0cm
14.0cm
6.0cm
14.0cm
Working Frequency
H6.0
3.5
H6.0
H6.0
PRF
1000
1000
1000
1000

Table of Contents

Related product manuals