C-18 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 35C50EA
Imaging Mode: B+M-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.98
0.52
1.04
1.70
Index component value
0.46
0.52
0.33
1.04
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
1.54
P (mW)
162.32
121.17
189.25
P
1
(mW)
29.12
28.65
z
s
(cm)
2.89
z
b
(cm)
4.91
z
MI
(cm)
4.91
z
pii
,α
(cm)
4.91
f
awf
(MHz)
2.46
3.36
2.46
2.46
Other
Informatio
n
prr (Hz)
1,000.0
0
srr (Hz)
37.00
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
212.55
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
172.25
I
spta
at z
pii
or z
sii
(mW/cm
2
)
398.34
p
r
at z
pii
(MPa)
2.23
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
14.8cm
14.8cm
14.8cm
14.8cm
B/M Focus position
6.0cm
14.0cm
6.0cm
14.0cm
B/M Working Frequency
H6.0
3.5
H6.0
H6.0
B PRF
3000
3000
3000
3000
M PRF
1000
1000
1000
1000