C-42 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 65EB10EA_S
Imaging Mode: Color+B-Mode / Power+B-Mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.27
0.16
0.11
0.29
Index component value
0.16
0.16
0.11
0.11
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
0.55
P (mW)
14.00
10.00
14.00
P
1
(mW)
6.25
4.46
z
s
(cm)
/
z
b
(cm)
/
z
MI
(cm)
0.98
z
pii
,α
(cm)
0.98
f
awf
(MHz)
4.06
5.40
5.65
4.53
Other
Information
prr (Hz)
2,332.00
srr (Hz)
13.00
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
9.44
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
0.34
I
spta
at z
pii
or z
sii
(mW/cm
2
)
0.45
p
r
at z
pii
(MPa)
0.63
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
8.3cm
8.3cm
8.3cm
8.3cm
B Focus Position
2.0cm
2.5cm
2.5cm
2.5cm
Color Sampling Gate
Position
2.0cm
2.5cm
2.5cm
2.5cm
B Working Frequency
5
6.5
6.5
5
C Working Frequency
4
5
5
4
B PRF
4889
3320
3320
2919
Color PRF
2332
4750
4750
4177

Table of Contents