EasyManua.ls Logo

Mindray DP-50 Vet - Page 300

Mindray DP-50 Vet
343 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C-52 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 65EL60EA
Imaging Mode: B-mode/Smart3D/Tissue Harmonic Imaging
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
1.23
0.04
0.10
0.03
Index component value
0.04
0.04
0.10
0.04
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
2.65
P (mW)
8.00
8.00
2.00
P
1
(mW)
1.32
1.32
z
s
(cm)
/
z
b
(cm)
/
z
MI
(cm)
1.56
z
pii
,α
(cm)
1.56
f
awf
(MHz)
4.65
5.99
5.87
4.43
Other
Informatio
n
prr (Hz)
5,096.0
0
srr (Hz)
13.00
n
pps
1.33
I
pa,α
at z
pii
,α
(W/cm
2
)
343.78
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
5.24
I
spta
at z
pii
or z
sii
(mW/cm
2
)
8.64
p
r
at z
pii
(MPa)
3.40
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
12.9cm
12.9cm
12.9cm
12.9cm
Focus position
1.5cm
6.0cm
6.0cm
5.0cm
Working Frequency
H9.0
6.5
6.5
H9.0
PRF
5096
5096
5096
5096

Table of Contents