C-62 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 75L38EA
Imaging Mode: B+PW-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
1.33
0.69
0.95
1.01
Index component value
0.69
0.58
0.48
0.95
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
3.17
P (mW)
47.14
24.21
45.96
P
1
(mW)
24.44
19.77
z
s
(cm)
1.67
z
b
(cm)
2.56
z
MI
(cm)
1.27
z
pii
,α
(cm)
1.27
f
awf
(MHz)
5.71
7.07
5.55
7.24
Other
Informatio
n
prr (Hz)
699.00
srr (Hz)
46.00
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
561.33
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
254.96
I
spta
at z
pii
or z
sii
(mW/cm
2
)
420.43
p
r
at z
pii
(MPa)
4.07
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
6.5cm
12.9cm
12.9cm
12.9cm
B Focus Position
1.0cm
10.0cm
3.0cm
10.0cm
SV Position
1.0cm
10.0cm
3.0cm
10.0cm
B Working Frequency
7.5
7.5
5
7.5
PW Working Frequency
5.7
5.7
5
5.7
B PRF
7694
1276
1276
1276
PW PRF
699
3594
3594
3594
PW SV
0.5mm
0.5mm
0.5mm
0.5mm