Acoustic Output Reporting Table (60601-2-37) C-91
Transducer Model: 75L50EAV
Imaging Mode: Color+B-Mode / Power+B-Mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.98
0.17
0.17
0.40
Index component value
0.17
0.17
0.17
0.17
Acoustic
Parameter
s
p
r,α
at z
MI
(MPa)
2.31
P (mW)
28.00
28.00
28.00
P
1
(mW)
5.83
5.83
z
s
(cm)
/
z
b
(cm)
/
z
MI
(cm)
1.18
z
pii
,α
(cm)
1.18
f
awf
(MHz)
5.59
6.81
6.81
5.93
Other
Information
prr (Hz)
2,642.00
srr (Hz)
47.00
n
pps
/
I
pa,α
at z
pii
,α
(W/cm
2
)
209.16
I
spta,α
at z
pii
,α
or z
sii
,α
(mW/cm
2
)
18.95
I
spta
at z
pii
or z
sii
(mW/cm
2
)
29.90
p
r
at z
pii
(MPa)
2.90
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
16.6cm
16.6cm
16.6cm
16.6cm
B Focus Position
1.5cm
8.0cm
8.0cm
7.0cm
Color Sampling Gate
Position
1.5cm
8.0cm
8.0cm
7.0cm
B Working Frequency
5
5
5
5
C Working Frequency
5
5
5
5
B PRF
802
1605
1605
1605
Color PRF
2642
3498
3498
3498

Table of Contents