16 - 44 Operator’s Manual
16 Probes and Biopsy
b. Insert a biopsy needle with the same specification as that of the guiding block into the
guiding block hole
4. Remove the needle-guided bracket:
a. Hold the probe and remove the needle.
b. Loosen the guiding block nut, and remove it from the guiding block fixing frame.
c. Rotate the clamping nuts of the needle guided-bracket. The needle guided-bracket is
separate from the probe. Hold the probe and take out the bracket.