16 Probes and Biopsy
Operator’s Manual 16 - 57
Compatible disinfectants for the magnetizer are shown in Table 16-7.
Cleaning
Perform the following procedure:
1. Wear sterile gloves to prevent infection.
2. Wash the magnetizer with clean water or soapy water to remove all the foreign matters, or,
wipe the magnetizer with a soft ethyl carbamate sponge.
Avoid using a brush, because it may damage the magnetizer.
3. Dry the magnetizer using a sterile cloth or gauze after rinsing. Do not dry the magnetizer by
heating it.
Disinfecting with Sprays or Wipes
CAUTION
Use protective eyewear when disinfecting using sprays.
Perform the following procedure:
1. Wear sterile gloves to prevent infection.
2. After you have finished cleaning, wipe or spray the magnetizer with a disinfectant.
Follow the disinfectant manufacturers recommended contact time and mode.
Table 16-7 Compatible disinfectants
Trade Name Manufacturer Chemical Name Procedures
Surfa’safe Anios Laboratoires didecyldimethylammonium chloride
1.4mg/g;
polyhexamethylene biguanide
hydrochloride 0.96mg/g
Sprays, fungicidal
in 20 min
T.B in 15 min
Sani-Cloth
®
Plus PDI n-Alkyl dimethyl benzyl ammonium
chloride;
n-Alkyl ethylbenzyl ammonium
chloride
Wipes, 3 min
Oxivir
TM/MC
Tb Diversey 0.5% hydrogen peroxide Sprays, 10 min
CLEANISEPT
®
WIPES
DR.SCHUMACHE
R GMBH
0.25g
didecyldimethylammoniumchloride;
0.5g quaternary ammonium
compounds; benzyl-C12-16-
alkyldimethyl-; chlorides
Wipes, 2 min
mikrozid
®
AF
Wipes Jumbo
SCHÜLKE 25% ethanol; 35% propan-1-ol Wipes, 15 min
IODOCLEAN Anios Laboratoires sodium thiosulfate and excipients Sprays, 5 ~ 15
min
tristel duo TRISTEL
SOLUTIONS
chlorine dioxide Sprays, 1 min
75% alcohol / alcohol Solution, 3 min
iodophor / / Solution, 3 min